MTS FlexTest Models 200 User Manual
Page 277
Model 493.07 HPU Converter Box
MTS FlexTest® Models 40/60/100/200 Controller Hardware
Digital I/O and Transition Boards
277
The Model 493.07 Converter Box can be connected to other controllers (as
shown in the following table) using a “Y” cable.
Note
If the Series 494 Chassis is the only device connected to the HPU,
the jumper settings are not necessary.
Note
The Model 493.07 Converter Box cannot be used with both groups
of controllers at the same time. If this functionality is required, use
an HPU isolation box.
C
OMPATIBLE
WITH
J
UMPERS
Model 458.05/.10/.20/.40
Model 490.01
Model 497.05
Standard jumper setting:
E10–E11
install
E12–E13 install
E14–E15 install
E16–E17 install
E10–E13 remove
E14–E17 remove
Model 413.05
Model 436.11
Model 407.05
Jumper change required:
E10–E11 remove
E12–E13 remove
E14–E15 remove
E16–E17 remove
E10–E13 install
E14–E17 install