beautypg.com

MTS FlexTest Models 200 User Manual

Page 147

background image

Digital Universal Conditioner Mezzanine Cards

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

147

Excitation Failure
Interlocks

An over- or under-current condition generates a
system interlock.

AC/DC over current: 105 mA typical

AC/DC under current: 1–2 mA typical

Note

Undercurrent detection may not work
properly for AC transducers with a DC
resistance greater than 180 ohms. In this
instance, the detection circuit may
constantly report an undercurrent condition.
You may need to configure your controller
software to disable the Excitation Failure
interlock for that input.

Excitation Failure interlocks are available for
quarter-, half-, and full-bridge DC configurations and
for AC configurations.

Smart Transducer
Interface

IEEE 1451.4 Class 2

*. The DC gain accuracy specification includes the effects of linearity

and resolution.

†. DC mV/V (ratio-metric) accuracy specifications apply in 8-wire mode

only.

‡. The DC excitation accuracy and loading specifications apply in 8-

Wire Mode only.

§. The effects of cable length, transducer impedance, and associated

voltage drop are additional.

#. Excitation loading variability applies for resistive loading.

DUC Specifications (part 3 of 3)

P

ARAMETER

S

PECIFICATION