background image

Model 494.16 VD/DUC Mezzanine Card

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

163

4. Install the VD/DUC card(s) on the I/O carrier board.

5. (Optional) Install shunt-calibration-resistor plug assemblies on the front

panel of the I/O carrier board(s).

Note

If you use MTS TEDS modules or MTS transducers with integrated
shunt-calibration resistors, you must insert a jumper plug (MTS part
number 100-188-097) into the socket for each transducer input
where you will use the integrated shunt-calibration resistor.

6. Install the I/O carrier board(s).

A. Plug the board into the VME card cage.

B. Connect the transducer cables to the I/O carrier board.

C. Connect the valve cable to the I/O carrier board.

7. After you have installed all boards, apply power to the system and use

the hardware-mapping and control software to configure the
conditioner and the valve driver.

8. Allow the system to warm up for an hour and then perform valve-

balance, calibration, and tuning procedures.