beautypg.com

MTS FlexTest Models 200 User Manual

Page 253

background image

Model 493.74 Two-Station HSM Interface Board

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Digital I/O and Transition Boards

253

Cable specification

To maintain EMC compliance, J43 Interlock cables must comply with the
following specifications:

Connector–9-contact, type D, male EMI connector.

Backshell–EMI metallized plastic or metal.

Cable–AWG and number of conductors as required. Braided shield with
shield connected to the metallized backshell at the chassis.

Jumper plug

If connector J43 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug part number 100-007-948 or
jumper pins 1 and 2; 3 and 4; 5 and 6; 7 and 8.