Model 494.25 single duc card settings – MTS FlexTest Models 200 User Manual
Page 155

Model 494.25 Single DUC Card
MTS FlexTest® Models 40/60/100/200 Controller Hardware
Mezzanine Cards
155
Model 494.25 Single DUC Card Settings
Hardware
configuration
The conditioner on this card has a two-position switch that is used to set up
the following DC conditioner functions:
•
Switch A connects a bridge completion resistor to -FBR to complete a
quarter-bridge circuit.
•
Switch B provides a ground reference for an external transducer by
connecting -FBR to ground.
Software configuration
After you set up and install the hardware, you must configure your control
software settings for the conditioner.
Note
For detailed information on software configuration, calibration, and
tuning, see the control software and tuning/calibration
documentation.
1 2
Switch Settings
A = Bridge Completion Switch
The ON position connects a
user-installed bridge completion
resistor to complete a quarter-
bridge circuit.
B = Grounding Switch
The ON position connects -FBR
line to ground. (See the bridge
completion drawings for details.)
SW1