Section 4: Commissioning

10/08 99

51207 Revision 1

1. Determining an action
Determine the action for the rising edge of the reference signal under [Limit 1 on] from the following
list (here: Marker 1 is set when 900 kg are exceeded).
Accordingly, an action for [Limit 1 off] can be determined.

Function

SPM bit

-no action-

-----

No function

set marker 1

X64=1

Set marker 1

set marker 2

X65=1

Set marker 2

set marker 3

X66=1

Set marker 3

select net

X72=1

Select net

set zero

X112=1

Set zero

set tare

X113=1

Set tare

reset tare

X114=1

Reset tare

set test

X115=1

Activate the analog test

reset test

X116=1

Finish the analog test

reset PWF

X117=1

Reset power fail

set fixtare

X118=1

Set fixtare (use the value in address D31 as a tare value)

get fixtare

X119=1

Save gross value as fixtare in address D31

set print

X120=1

Activate print order ?

clr marker 1

X64=0

Clear marker 1

clr marker 2

X65=0

Clear marker 2

clr marker 3

X66=0

Clear marker 3

select gross

X72=0

Save the gross weight in address D11


2. Determining a condition
Additionally, a condition [Condition] can be assigned to the signal; see Chapter 4.8.4.

4.8.6

BCD Output


As a prerequisite for using the BCD output, a BCD card must be installed; see Chapter 3.4.6 and
Chapter 3.4.7.

Setup/Digital I/O parameter

Output 1

marker bit 1

X64=1

Output 2

marker bit 2

X65=1

Output 3

marker bit 3

X66=1

Input 1 on

-no action-

Input 1 off

-no action-

Input 2 on

-no action-

Input 2 off

-no action-

Input 3 on

-no action-

Input 3 off

-no action-

BCD out

Gross


[BCD
out]

Output as BCD with max. 5 decades with minus or plus sign and
status:

[Gross]

Gross weight

[Net if tared]

Net weight if tared, else gross weight

[Selected]

Gross or Net, depending on SPM-Bit X72

[Transparent]

The value in D20 is output as BCD