Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 99
Section 4: Commissioning
10/08 99
51207 Revision 1
1. Determining an action
Determine the action for the rising edge of the reference signal under [Limit 1 on] from the following
list (here: Marker 1 is set when 900 kg are exceeded).
Accordingly, an action for [Limit 1 off] can be determined.
Function
SPM bit
-no action-
-----
No function
set marker 1
X64=1
Set marker 1
set marker 2
X65=1
Set marker 2
set marker 3
X66=1
Set marker 3
select net
X72=1
Select net
set zero
X112=1
Set zero
set tare
X113=1
Set tare
reset tare
X114=1
Reset tare
set test
X115=1
Activate the analog test
reset test
X116=1
Finish the analog test
reset PWF
X117=1
Reset power fail
set fixtare
X118=1
Set fixtare (use the value in address D31 as a tare value)
get fixtare
X119=1
Save gross value as fixtare in address D31
set print
X120=1
Activate print order ?
clr marker 1
X64=0
Clear marker 1
clr marker 2
X65=0
Clear marker 2
clr marker 3
X66=0
Clear marker 3
select gross
X72=0
Save the gross weight in address D11
2. Determining a condition
Additionally, a condition [Condition] can be assigned to the signal; see Chapter 4.8.4.
4.8.6
BCD Output
As a prerequisite for using the BCD output, a BCD card must be installed; see Chapter 3.4.6 and
Chapter 3.4.7.
Setup/Digital I/O parameter
Output 1
marker bit 1
X64=1
Output 2
marker bit 2
X65=1
Output 3
marker bit 3
X66=1
Input 1 on
-no action-
Input 1 off
-no action-
Input 2 on
-no action-
Input 2 off
-no action-
Input 3 on
-no action-
Input 3 off
-no action-
BCD out
Gross
[BCD
out]
Output as BCD with max. 5 decades with minus or plus sign and
status:
[Gross]
Gross weight
[Net if tared]
Net weight if tared, else gross weight
[Selected]
Gross or Net, depending on SPM-Bit X72
[Transparent]
The value in D20 is output as BCD