Section 4: Commissioning

10/08 62

51207 Revision 1

4.3.3 Parameter Table

- SEtuP

- Cd

Calibration data

- 000

CALIb

nEW new calibration, VIEW view data, Mod modify calibration

- 001

MAX.FSd

Decimal point position, Max (Scale range) and weight unit

- 002

StEP

Scale interval

- 003

dEAdLo

Deadload setting by weight (LoAd) or mV/V (MVoLt)

- 004

SPAn

Calibration weighing range by weight (LoAd) or mV/V (MVoLt)

- 005

0-rnG

Show used zeroset range

- 006

uVolt.d

Show µV/d

- 007

WEIGHt

Show current gross weight, with

10-fold multiplier is activated

- CP Calibration parameter

- 010

MEAtIM

Measure time

- 011

FILtEr

Digital filter

- 012

FCut

Cut-off frequency

- 013

tStMod

Test mode

- 014

WAM

W&M, legal for trade

- 015

StStIM

Standstill time

- 016

StSrnG

Sandstill range

- 017

TArtIM

Time out for taring / zeroset

- 018

ZESrnG

Zeroset range

- 019

ZEtrnG

Zerotrack range

- 020

ZEtStP

Zerotrack step

- 021

ZEttIM

Zerotrack time

- 022

oVrLd

Overload

- 023

MIn

Min weight

- 024

MuLrnG

Multi range mode

- 025

rAnG 1

Range 1

- 026

rAnG 2

Range 2

- LI

Limits

- 030

LIM.1on, LIM.1oF Limit 1 'On', Limit 1 'Off'

- 031

LIM.2on, LIM.2oF Limit 2 'On', Limit 2 'Off'

- 032

LIM.3on, LIM.3oF Limit 3 'On', Limit 3 'Off'

- LA

Limits, action / condition

- 033

LIM.1on

Action / Condition

- 034

LIM.1oF

Action / Condition

- 035

LIM.2on

Action / Condition

- 036

LIM.2oF

Action / Condition

- 037

LIM.3on

Action / Condition

- 038

LIM.3oF

Action / Condition

- do

Digital outputs

- 040

bCd.out

BCD-Mode: GroSS = gross, NEtIt = net if tared, else gross, SELECt,
trnSP

- 041

outP. 1

Output 1

- 042

outP. 2

Output 2

- 043

outP. 3

Output 3

- dI

Digital inputs

- 044

InP.1on

Input 1, condition

- 045

InP.1oF

Input 1, condition

- 046

InP.2on

Input 2, condition

- 047

InP.2oF

Input 2, condition

- 048

InP.3on

Input 3, condition

- 049

InP.3oF

Input 3, condition

- Ao

Analog outputs

- 050

AnA.Mod

Mode: GroSS = Brutto, NEtIt = Netto wenn tariert, sonst Brutto, SELECt,
trnSP

- 051

AnA.rnG

Range 0/4 - 20 mA

- 052

out.Err

Output at error

- 053

out.< 0

Output at < 0

- 054

out.>20

Output at > Max

- 055

WGt.0/4

Weight at 0/4 mA