beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 121

background image

Section 5: SMA Protocol

10/08 121

51207 Revision 1

Request for Taring with Fixtare Value

Command: T
Reply:

The scale makes a taring attempt using the fixtare value and signals the
tared status in the and characters.

For details, see Chapter 5.4.1 .



Request for Zero Setting of the Scale

Command: Z
Reply:

The scale makes a zero setting attempt and signals the zero status in the
character.

For this function, the stability condition must be met. The maximum waiting time for
stability is set under [Tare timeout]; see Chapter 4.5.13 .

For details, see Chapter 5.4.1 .

Unless the scale is in the zero setting range, an error reply is generated.



Request for Tare Resetting

Command: C
Reply:

The scale deletes the tare weight and signals the tare reset status in the status
character. The scale tare is reset.

For details, see Chapter 5.4.1 .


5.3.3 Scale Diagnosis


Command: D
Reply:

The scale starts the diagnosis and returns a diagnosis reply.

For details, see Chapter 5.4.1 .