Section 4: Commissioning
10/08 93
51207 Revision 1
4.8 L
IMIT
V
ALUES
,
D
IGITAL
I
NPUTS AND
O
UTPUTS
4.8.1 Conditions for Limit Values and Digital Inputs, States for Outputs
The limit values and digital inputs can be combined with conditions that must be met before an
action is executed. With outputs, these conditions can be output as states directly. Selection is from
the following list:
Selection list for [condition]
Function SPM bit (dec)
Explanation
Condition is met, if not true (=0) or if true (=1)
no condition
-----
No condition
actual diginp1
X00=0/1
Digital input 1
actual diginp2
X01=0/1
Digital input 2
actual diginp3
X02=0/1
Digital input 3
actual limit 1
X16=0/1
Limit signal 1
actual limit 2
X17=0/1
Limit signal 2
actual limit 3
X18=0/1
Limit signal 3
ADC error
X32=0/1
Error in load cell/ ADC area, negative/ far outside scale
above Max
X33=0/1
Weight above Max
overload
X34=0/1
Weight above (Max plus the range which is still permissible)
below zero
X35=0/1
Weight below zero
center zero
X36=0/1
Within 1/4 d of zero
inside ZSR
X37=0/1
Within zero-setting range
standstill
X38=0/1
Mechanical stability of the scale
dimmed
X39=0/1
Weight below zero or above Max
command error
X48=0/1
Command was no executed; e.g. zero-setting command, but
the weight is out of the zero-setting range
command busy
X49=0/1
Command is being executed
power fail
X50=0/1
Set after power-on (=power failure)
test active
X56=0/1
Analog test was started
cal active
X57=0/1
Calibration was started
tare active
X58=0/1
Tare step is active
marker bit 1
X64=0/1
Marker bit 1; after power-on, the markers are set to '0'
marker bit 2
X65=0/1
Marker bit 2,
marker bit 3
X66=0/1
Marker bit 3