Section 3: Installing the Instrument and Plug-In Cards

10/08 21

51207 Revision 1

Connecting a Remote Display / Remote Terminal

A PR 1627 remote display or a PR 1628 remote terminal can be connected to the built-in RS-232
interface [Builtin RS232] or to the PR 5510/04 card.

Press

-[Serial ports parameters]-[Remote display]-[Builtin RS232]-[Param] and select

[Baudrate] 4800. The baud rate now corresponds to the default settings of PR 1627 or PR 1628.
The following settings cannot be changed: [Bits] 7, [Parity] even and [Stopbits] 1.
If only one instrument is connected to a PR1627/8, [Mode] must be [single transmitter].
When connecting several instruments to a PR 1628 via an RS-232/RS-485 converter, selection for
display on PR 1628 is possible using addresses. For this purpose, select [multiple transmitters] as
[Mode] in all instruments, enter the instrument address under [Device Id] and the address of the
subsequent instrument under [Next Device Id].


Connecting a YDP12IS or YDP04IS Ticket Printer

The ticket printer YDP12IS-OCEUV or YDP04IS-OCEUV can be connected via [Builtin RS232]
interface or the RS-232 on card PR 5510/04.


If the printer is connected to the [Builtin RS232] port:

Press

-[Serial ports parameter]-[Printer]-[Builtin RS232]-[Param] and configure the following

settings under [Protocol]: [RTS/CTS], [Baudrate]: 9600, [Bits]: 8, [Parity]: [none], [Stopbits]: 1 and
[Output mode]: [Raw].
The printer must be set to Line Mode (factory setting: Page Mode). Press the 'FEED' button to
change modes; please refer to the installation instructions delivered with the printer.