Section 3: Installing the Instrument and Plug-In Cards

10/08 25

51207 Revision 1

3.4.2 Cable Connection in the D-Sub Connector Mating Plug


The connections on the back panel are plug-in type. Keep the conductors as short as possible and
connect them to the terminals. The connector housings are conductive (metallized), i.e., part of the
shield, and must be fastened to the back panel by screws.

Mounting a cable:

Open the connector housing (catches)

Release and open the cable clamp

Remove approx. 50-60 mm of the cable insulation

Shorten the shield to 5 mm and bend it over the cable

sheath

Remove 3mm wire insulation and connect it by soldering

Insert the pin unit

Put the cable under the cable clamp

The grounding tongue presses on the shield bent
backwards; the clamp presses on the cable sheath

Close and tighten the cable clamp

Check the strain relief

Insert the mounting screws on both sides

Close the connector housing (catches)


The shields must be connected to the metal housings on
both ends of the cable.