Section 4: Commissioning
10/08 97
51207 Revision 1
4.8.5 Configuring Limit Values
Each limit value consists of a switch-on and a switch-off point for definition of a hysteresis. The
three pairs of values must be entered according to the same principle. The limit values always refer
to the gross weight. Values between -0.01 x Max and 1.01 x Max of the relevant scale are
permissible.
Setup
Serial ports parameter
Date & Time
Operating parameter
Printing parameter
Fieldbus parameter
Network parameter
Calibration
Limit parameter
Press
and
Digital I/O parameter
to select [Limit parameter]
Setup/Limit parameter
Determine the limit values
Limit 1 on
890 kg
Action
-no action-
Limit 1 off
900 kg
Action
-no action-
Limit 2 on
300 kg
Action
-no action-
Limit 2 off
290 kg
Action
-no action-
Setup/Digital I/O parameter
Assign the outputs
Output 1
limit 1 out
X16=
1
Output 2
limit 2 out
X17=
1
Example: