Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 75
Section 4: Commissioning
10/08 75
51207 Revision 1
4.5.4 Determining the Scale Interval
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1kg The scale interval
1, 2, 5 10, 20 or 50 can be set by pressing
.
The weight unit is taken from [Max] and cannot be changed here.
The number of digits behind the decimal point must be determined already when entering [Max] as
well.
Set Scale interval
After
or
confirmation of the change is displayed with:
Set scale interval failed
Max not multiple of scale interval
The maximum capacity is not an integer multiple of the scale
interval.
4.5.5 Determining the Dead Load
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Not calibrated
Sensitivity
833.33
4.000000
by load
by mV/V
Test
To use the empty scale/hopper as dead load (normal case):
- discharge the scale/hopper
- press [by load]
Set deadload
After
or
confirmation of the change is displayed:
If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the
value can be overwritten by pressing [by mV/V].
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
Test