Section 4: Commissioning

10/08 75

51207 Revision 1

4.5.4 Determining the Scale Interval

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1kg The scale interval

1, 2, 5 10, 20 or 50 can be set by pressing

.


The weight unit is taken from [Max] and cannot be changed here.
The number of digits behind the decimal point must be determined already when entering [Max] as
well.

Set Scale interval

After

or

confirmation of the change is displayed with:

Set scale interval failed

Max not multiple of scale interval

The maximum capacity is not an integer multiple of the scale
interval.

4.5.5 Determining the Dead Load

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

0.00 kg

0.000000

Max at

3000.00 kg

1.000000

Not calibrated

Sensitivity

833.33

4.000000

by load

by mV/V

Test


To use the empty scale/hopper as dead load (normal case):
- discharge the scale/hopper
- press [by load]

Set deadload

After

or

confirmation of the change is displayed:


If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the
value can be overwritten by pressing [by mV/V].

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

0.00 kg

0.000000

Max at

3000.00 kg

1.000000

Calibrated at

3000.00 kg

1.000000

Sensitivity

833.33

4.000000

by load

by mV/V

Test