Section 3: Installing the Instrument and Plug-In Cards
10/08 36
51207 Revision 1
3.4.7 PR 5510/09 BCD Output (Open Collector)
The plug-in card is used for BCD-coded output of a 5-decade weight value.
Cyclical output of the weight value is with 4-bit corresponding status (polarity sign, stability, error).
The value is intrinsically consistent and can be output continuously or the digital input DATA_IN
control can be used to freeze (hold) the value.
Internal connection:
34-contact female connector on
flat cable for Slot 1, 2
Number of outputs:
5 digits BCD + plus or minus sign
Number of inputs:
Input: 1 bit (DATA IN)
Output stage:
Common emitter at GND,
open collector
External supply:
+5 V ... +24 VDC
Voltage drop:
Approx. 0.9 V
Output current:
Max. 50 mA
Enable input:
5 V / 24 V adjustable via S101
@ 5 V high > 3.1 V, low < 1.5 V
@ 24 V high > 16 V, low < 10V
protected against incorrect polarity
Cable:
Shielded, max. 50 m
Electrical isolation:
No
External connection:
X102
26-contact D-Sub connector
(female)
Dimensions:
(LxWxH):
60x106x22
Accessories (delivered
with the unit):
26-pin D-Sub mating connector
(male)
incl. shielded housing
Weight:
55 g
Before installation, set the switches S101 and S102 as given in Chapter 0.