Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 79
Section 4: Commissioning
10/08 79
51207 Revision 1
[Nominal load]
Nominal load E
max
of a load cell (not the overall nominal weight of the scale).
[Gravity]
Acceleration of gravity at the place of installation; default is the value for Hamburg, Germany,
9.81379 m/s².
[Hysteresis error]
When switching from [Not specified] to [Specified], values for [Correction A/B] must be filled in. The
data is given on the load cell certificate.
[Certified data], [LC sensitivity], [LC resistance]
With [all LC same] only 1 value for the sensitivity [LC sensitivity] and the output resistance
[LC resistance] must be filled in. With [each LC specific] individual values for each load cell are
requested.
[Calc]
The mV/V value is calculated and after confirmation with [OK], the calculated mV/V value is stored
in the calibration data.
4.5.9 Subsequent Dead Load Correction
If the hopper/platform weight changes by an amount that is higher than the zero-setting range; e.g.,
due to dead load reduction, dead load increase or mechanical changes, the functions for automatic
zero tracking and manual zero setting no longer work. To view the range which is already utilized by
zero tracking or zero setting, in [Calibration] press the
key; this also activates 10-fold increased
resolution of the weight value. Press
again to return to the previous state:
Current zero setting: 0.123 kg
If the entire zero-setting range is already utilized, you can still correct the dead load (CAL switch
must be open) without affecting other calibration data/parameters.
To do this, select
-[Calibration]-[Modify] and determine the dead load with [Deadload at]-[by
load] (see Chapter 0).