Section 4: Commissioning

10/08 58

51207 Revision 1

4.2 S

WITCHING ON THE

I

NSTRUMENT

The instrument can be put into operation and calibrated using a PC with the VNC program (on the
enclosed CD), an Internet browser or the ConfigureIt program.
On initial start-up, the date and time must be set.
When the supply voltage is applied to the instrument, the following information is displayed:

xPR5410w

Instrument type: PR5410 = X3

xB,10,00,0w

BIOS release

xF,10,00,0w

Firmware release

z,8,8,8,8,8,8y

Automatic display test

m135,050b

Weight display

xERROR6

Error message if no load cells are connected, see Chapter 4.6.1.


4.2.1 Display Test

Press

to access the front panel key menu:

xDS,PTS,Tx

.

z,8,8,8,8,8,8y

Press

to activate all display items. After approx. 5 s,

the menu is closed automatically.


4.2.2 Front-Panel Key Test

Press and hold

, then press

as well.

x,0,0,0,0,0,0w

One digit per key is displayed. Each time a key is pressed, the corresponding digit is
incremented by 1

x,0,0,0,1,0,0w

Display, if

is pressed once

x-8023-w

Display, if several keys (2 and 3) are triggered simultaneously


If no key is pressed for 5 s, the test is finished automatically and the weight is displayed again.

4.2.3 Resetting the Instrument to the Factory Settings


Reset to the factory settings is possible only, when the CAL switch is open.

Press Setup at the instrument

DS,PTS,Tw

Press

to reach the BIOS menu

BIOS w

Press

BIO,S..w

Is displayed, followed by

FLASH w

Press

to go to menu

DE,LEA,Rw

Press

ERA,S 0w

It is asked, whether resetting should be done

ERA,S 1w

Alter with

the '0' to '1' for resetting and confirm with

DONE w

Is displayed to indicate the execution