Section 4: Commissioning

10/08 84

51207 Revision 1


Multi-interval scale
With [Range mode] = [Multi-interval], the scale has up to three ranges with different resolution. Each
range has the corresponding interval. Unlike [Multiple range], switching the interval is also triggered
by weight reduction; i.e., when the weight drops below the range limits.

During calibration, the multiple range/multi-interval function is always switched off.
The weight display header (VNC) includes the current range (R1, R2, R3), Max, Min and d (or e with
legal-for-trade instruments) (Example: multiple range scale in range 2):


The ranges are marked by points on the left side of the display.

Range 1

Range 2

Range 3