Section 4: Commissioning

10/08 85

51207 Revision 1

4.6 E

RROR

M

ESSAGES

4.6.1 Measuring Circuit Error Messages

The weighing electronics can generate error messages, which are output on the weight display.

Display

VNC text

Cause

EPPOP1

Arith err

Internal arithmetic overflow (faulty calibration values)

EPPOP2 Overload

Input voltage is higher than Max + (x d)

EPPOP3 No EOC

Input signal is higher than the permissible range of 36 mV.
However, the trouble cause can also be due to an error in the
analog section, to a defective load cell or to a cable break

EPPOP6 No sense

voltage

Load cells not connected. Sense line or supply line is interrupted or
connected with wrong polarity or sense voltage is low

EPPOP7 Negative Input

Input voltage is negative (e.g. wrong polarity of load cell signal or
supply voltage)


Testing the Measuring Circuit
A simple test with the load cells connected can be carried out with a multimeter
(not with external supply or intrinsically safe load cell interface):

12 VDC +/- 0.8 V
(symmetrical to housing GND)

12 VDC +/- 0.8 V
(symmetrical to housing GND)

0 - 12 mVDC @ LC with 1.0
mV/V
0 - 24 mVDC @ LC with 2.0
mV/V

Supply voltage

Sense voltage

Measuring voltage

4.6.2 General Error Messages

Display

Cause

EPP 21

The BIOS is too old for the firmware

EPP 22

The firmware is too old for the BIOS

EPP 23

Starting is not possible; ROM error

EPP 25

The plug-in card combination is not permissible

EPP 29

Flashing failed

EPP 31

The IP address is faulty, e.g. item > 255

EPP 32

The IP address is already used

EPP 33

Wrong parameter

EPP 41

Fatal system error

EPP 42

Fatal control system error