Section 7: Fieldbus Interface

10/08 137

51207 Revision 1

7.3.6 Register 5: Device Type and Software Release


Only reading is permitted.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Byte 0

TYPE MSB

Byte 1

TYPE LSB

Byte 2

MAINVERSION

Byte 3

SUBVERSION


E.g. 5410 Rel 1.23 = 54100123

hex


7.3.7 Register 6: Board Number

Only reading is permitted.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Byte 0

Board number MSB

Byte 1

""

Byte 2

""

Byte 3

Board number LSB


E.g. 148388723 = 08D83B73

hex


7.3.8 Register 7: (Reserved)


7.3.9 Register 8 ...15: Weight Data


Only reading is permitted.
The gross, net and tare weight are stored as DINT fixpoint.
The real data value is derived from DINT and EXPO as follows :

Value

Real

=reading

DINT

*10

(-EXPO

Register 8

Current gross value

Register 9

Current net value, if tared; otherwise gross

Register10

Current tare value, if tared, otherwise 0

Register11

Value on the front-panel display

Register12

Reserved

Register13

Reserved

Register14

Max

Register15

Reserved

7.3.10 Register 20: Weight Data

Register 20

BCD output value


7.3.11 Register 22 ... 27: Limit Values (Read/Write)

Register 22

Limit 1 on

Register 23

Limit 1 off

Register 24

Limit 2 on

Register 25

Limit 2 off

Register 26

Limit 3 on

Register 27

Limit 3 off