Section 3: Installing the Instrument and Plug-In Cards

10/08 53

51207 Revision 1

3.5.3 Connecting up to 8 Load Cells (650 Ohms) Using 6-Wire Connecting Cable


Via cable junction box PR6130/.. using PR 6135 or PR6136 connecting cable:

The cable colors
shown here are valid
for PR 62 . . series
load cells and for the
PR 6135 connecting
cable.

When using different
load cells, the cable
colors may have
different
assignments.

For this reason,
consult the relevant
manual or data sheet
for the assignments
of cable colors
before connecting.



Recommendations: - Install cable in steel pipe connected to earth potential.

- Min. distance to high-voltage cables: 1m


Load cell supply circuit:

Load resistance of load cell circuit > 75 ohms, e.g. 8 load cells of 650 ohms each,

supply voltage is 12 V DC; for further data, see Chapter 12.4.