beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 125

background image

Section 5: SMA Protocol

10/08 125

51207 Revision 1

5.4.5 Reply with ‘A’ and ‘B’ Command


Reply format with ‘A' and 'B' commands (variable length):
:
The reply format and meaning are:

Start of reply from 'A'/'B' command

The field name is three characters long, left-adjusted and filled up with spaces on
the right, if necessary.

The following fields are sent:

“SMA” level/revision

(reply from ‘A’ command)

“MFG” manufacturer marking

(reply from the 1st ‘B’ command)

“MOD” product / model identification

(reply from the 2nd ‘B’ command)

“REV” software version

(reply from the 3rd ‘B’ command)

“SN_“ serial number

(reply from the 4th ‘B’ command)

“END”: This is always the last inscription field

(reply from the last ‘B’ command)


‘:’

Separator between field name and field content.

The data field contains up to 25 character.

The SMA field contains

with the following meaning: level = (1, 2, etc.); revision = (1.0; 1.1; etc.)

End of reply from 'A'/'B' command



Example:
Command

Reply

A

SMA:1/1.0

B

MFG:Sartorius

B

MOD:PR5410

B

REV:01.01.9

B

SN_:148388723

B

END:


Note: If the controlling computer sends another 'B' command:
B ?