Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 24
Section 3: Installing the Instrument and Plug-In Cards
10/08 24
51207 Revision 1
3.4 A
CCESSORIES
3.4.1 Installing Plug-in Cards
The main board has two slots with identical pin allocation (34 contact pins in two rows of 17) and
another slot (34 contact pins in two rows of 17) for plug-in cards. The slot designations are “Slot 1 ...
2" and "Slot 4" (left). Up to 2 cards can be mounted. Accordingly, the back panel is provided with
two cut-outs for the retainer plates of the cards.
Before installing or removing a plug-in card, disconnect the instrument from all voltage
sources.
View from above, back panel connectors at the
bottom
Installation (Slots 1-2, 4):
The flat cables plug into connectors (Slots 1
... 2, 4) on the main board. The cables are
polarized; i.e. incorrect polarity is precluded.
A description and examples of the various
cards and connections are given in Chapter
3.3.3 .
Remove the dummies from the back panel (2
screws; M3) and replace them with the
retainer plates for the plug-in cards.
The flat-cable connectors must be inserted
into the corresponding slots on the main
board.
View from the back
After installation/modification, the plug-in cards are detected automatically.
To view a list of the installed plug-in cards, select
-[Show HW-slots]:
Info/HW-Slots
Builtin
RS 232 Built-in serial interface
Slot 1
PR
5510/04
RS 485/232 Slots (Slot 1 and 2) are identical
Slot 2
-empty-
Slot 3
Builtin
Digital I/O Built-in digital I/Os
Slot 4
PR
1721/31
Profibus-DP
Only Fieldbus cards