Section 4: Commissioning

10/08 73

51207 Revision 1

4.5.2 Selecting the Calibration Mode

You can choose between [New] and [Modify] with the softkeys:

New

Modify

Param



New Calibration

Open the menu via

-[Calibration].

When you press [New], the data is set to default first and calibration is started.

SPAN and deadload

will be reset

You are prompted to confirm:
Press [Continue] for the default settings, or [Cancel] to cancel the
selection.


Default settings with [New]:

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

0.000000

Max at

1.000000

Not calibrated

Sensitivity

833.33

4.000000

Test


Changing the Calibration

[Modify] may be used only for minor changes (e.g. changing the dead load, adapting mV/V
values for dead load and/or Max); otherwise, always use [New].

Open the menu via

-[Calibration]-[Modify].

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

1.07 kg

0.000358

Max at

3000.00 kg

1.000000

Calibrated at

3000.00 kg

1.000000

Sensitivity

833.33

4.000000

by load

by mV/V

Test


For setting a new value for Dead load, press

to select [Deadload] and either enter a new

value with [by mV/V] or discharge the scale/hopper and press [by load].

Exit calibration

without CalcTest?

When closing the menu with

you are prompted whether the menu

should be closed without calculation of the test value:
Reply [Yes] to close the menu.