Section 4: Commissioning
10/08 83
51207 Revision 1
[Overload]
Weighing range above the maximum capacity (Max), without error message. Available range: 0 to
999900 d. In 'legal-for-trade' mode max. 9 d=e has to be entered.
WP A/Calibration
Min
50 d
Range mode
Single range
[Min]
Minimum weight at which a print command can still be executed. Range is 0 to 999900 d.
In 'legal-for-trade' mode min. 20 d has to be entered.
Legal-for-Trade Operation
Under
-[Calibration]-[Param]-[W
&
M] you can choose between [none] and a legal-for-trade mode:
[OIML], [NTEP] or [NSC].
[none]
[OIML]
[NTEP]
[NSC]
Gross weight display
0.125 mV/V
at 30000 d
0.125 mV/V
at 3000 e
0.125 mV/V
at 3000 e
0.125 mV/V
at 3000 e
0.25 mV/V
at 60000 d
0.25 mV/V
at 6000 e
0.25 mV/V
at 6000 e
0.25 mV/V
at 6000 e
Min. meas. signal
0.42 mV/V
at 10000 e
0.42 mV/V
at 10000 e
0.42 mV/V
at 10000 e
If legal-for-trade operation is selected, the parameters (zero tracking etc.) must be set accordingly;
they are not checked. The relevant CAL switches (see Chapter 4.1.1) must be sealed in the closed
position.
Multiple Range Scale/Multi-Interval Scale
Range selection is controlled by three parameters under
-[Calibration]-[Param].
WP A/Calibration
Range mode
Multiple range Select [Multiple range] or [Multi-interval]
Range limit 1
1000 kg Switch point from range 1 to 2
Range limit 2
2000 kg Switch point from range 2 to 3
Multiple range scale
With [Range mode] = [Multiple range], the scale has up to 3 ranges with different resolutions.
The switch points [Range limit 1] and [Range limit 2] are the range limits. As soon as the gross
weight exceeds range 1, the next higher range with the next higher interval becomes valid (1->2->5-
>10->20->50->100->200). When reducing the weight, the interval of the previous range is kept.
When the gross weight is <= 0.25 d of range 1 and the scale is stable and not tared, the scale
returns to range 1 with the corresponding interval.