Section 4: Commissioning

10/08 72

51207 Revision 1

4.5 C

ALIBRATION

Calibration using weights, mV/V or load cell data can be done using the VNC program. During
calibration, the instrument must be set to gross weight display (reset tare, if necessary).

For a ’legal-for-trade’ application, set the mode under

-[Calibration]-[Param] to [W&M] before

starting calibration; see Chapter 0. Select [New] to go to the maximum capacity [Max] (see Chapter
4.5.3); select the scale interval and determine the dead load. Now calibrate the maximum capacity
by applying a calibration weight, in mV/V or with load cell data.
After determining the test value, the menu can be closed as described in Chapter 4.5.12, in order to
save the new settings. Calibration data can be protected by the CAL switch (see Chapter 4.1.1),
which must be sealed in the closed (write-protected) position for ‘legal-for-trade’ applications.

4.5.1 Displaying Calibration Data

Cannot calibrate!

CAL_switch is closed

When the CAL switch is closed, the following message is displayed,
only data display possible with [Param]:

Cannot calibrate!

Scale is tared

If the scale was tared, the following message is displayed, data
display with [View], reset tare with [Res.tar.], return with [Cancel]:


With [View], the calibration data can be displayed, but not changed.

WP A/View Calibration

Max 3000 d 3000 kg

Number of scale intervals and max.
capacity

Scale interval 3000 d 1 kg

Scale interval

Deadload at 0.00 kg 0.000000

Dead load in weight and mV/V

Max at 3000.00 kg 1.000000

Weight and mV/V for maximum capacity

Calibrated at 3000.00 kg 1.000000

Test load* and corresponding mV/V

Sensitivity 833.33

4.000000

Number of internal counts and voltage

per scale interval

Param

The calibration data and parameters (press [Param]) are displayed in the format entered/determined
during calibration.
* After input with mV/V, the maximum capacity and the mV/V value entered are displayed.


Increased Resolution (10-Fold)

In the

-[Calibration] menu, the weight is displayed with 10-fold resolution (also with the CAL

switch closed) when you press the key

, and marked as an invalid weight with

above the

weight unit. The display on the instrument remains unchanged. After 5 s, the VNC display returns to
normal resolution, or you can press the key

to return to normal display immediately.