Section 4: Commissioning
10/08 72
51207 Revision 1
4.5 C
ALIBRATION
Calibration using weights, mV/V or load cell data can be done using the VNC program. During
calibration, the instrument must be set to gross weight display (reset tare, if necessary).
For a ’legal-for-trade’ application, set the mode under
-[Calibration]-[Param] to [W&M] before
starting calibration; see Chapter 0. Select [New] to go to the maximum capacity [Max] (see Chapter
4.5.3); select the scale interval and determine the dead load. Now calibrate the maximum capacity
by applying a calibration weight, in mV/V or with load cell data.
After determining the test value, the menu can be closed as described in Chapter 4.5.12, in order to
save the new settings. Calibration data can be protected by the CAL switch (see Chapter 4.1.1),
which must be sealed in the closed (write-protected) position for ‘legal-for-trade’ applications.
4.5.1 Displaying Calibration Data
Cannot calibrate!
CAL_switch is closed
When the CAL switch is closed, the following message is displayed,
only data display possible with [Param]:
Cannot calibrate!
Scale is tared
If the scale was tared, the following message is displayed, data
display with [View], reset tare with [Res.tar.], return with [Cancel]:
With [View], the calibration data can be displayed, but not changed.
WP A/View Calibration
Max 3000 d 3000 kg
Number of scale intervals and max.
capacity
Scale interval 3000 d 1 kg
Scale interval
Deadload at 0.00 kg 0.000000
Dead load in weight and mV/V
Max at 3000.00 kg 1.000000
Weight and mV/V for maximum capacity
Calibrated at 3000.00 kg 1.000000
Test load* and corresponding mV/V
Sensitivity 833.33
4.000000
Number of internal counts and voltage
per scale interval
Param
The calibration data and parameters (press [Param]) are displayed in the format entered/determined
during calibration.
* After input with mV/V, the maximum capacity and the mV/V value entered are displayed.
Increased Resolution (10-Fold)
In the
-[Calibration] menu, the weight is displayed with 10-fold resolution (also with the CAL
switch closed) when you press the key
, and marked as an invalid weight with
above the
weight unit. The display on the instrument remains unchanged. After 5 s, the VNC display returns to
normal resolution, or you can press the key
to return to normal display immediately.