beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 76

background image

Section 4: Commissioning

10/08 76

51207 Revision 1

Possible error messages:

Set deadload failed

above physmax

The dead load entered in mV/V plus maximum capacity in mV/V is
higher than 3 mV/V ( = 36 mV).

Set deadload failed

no standstill

The scale is not stable.
Remedial action: Check the mechanical function of the scale; adapt the
filter setting; reduce the resolution; if necessary, adapt the stability
conditions.

Set deadload failed

deadload < -0.1mV/V

Measuretime signal is negative (load cells connected with wrong
polarity or defective) when determining the dead load with [by load].

Set deadload failed

overflow in arithmetics

The dead load entered in mV/V is higher than 5mV/V.

4.5.6 Calibration with Weight (by Load)

Select [by load] for calibration using weight.

Place CAL weight

on the scale

and enter value

The weight value for the calibration weight must be entered in a
separate window.

2000

kg


After applying the weight, enter the weight value and confirm with

. The weight unit for the

calibration weight (press

to change) may differ from the unit in the instrument; conversion is

automatic.
Afterward, the following message is displayed:

Setting SPAN by load

Weight value, weight unit and measuring signal in mV/V corresponding to this value are displayed in
the [Calibrated at] line.

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

165.11 kg

0.057920

Max at

3000.00

kg

1.052369

Calibrated at

2000 kg

0.701579

Sensitivity

876.97

4.209600

by load

by mV/V

by data

Linear.

Test

Set SPAN failed

No stability

The scale is not stable.
Remedial action: Check the mechanical function of the scale; adapt
the filter setting; reduce the resolution; if necessary, adapt the
stability conditions.