beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 102

background image

Section 4: Commissioning

10/08 102

51207 Revision 1

4.10 A

LIBI

M

EMORY

The instrument is equipped with an alibi memory for approx. 80000 items. If the number is
exceeded, the eldest items are overwritten. No license is required for using the alibi memory. At
each print command a dataset with a sequence number is filed. The dataset can comprise of up to 3
items (Gross, Net and Tare).

Each item consists of :

- Sequence number

1 - 999999

- Date

YYYY-MM-DD

- Time

hh:mm:ss

- Scale address

Always A

- Weight type

B = Gross, N = Net, T = Tare, PT = Preset tare

- Weight with unit

Separation of decimals is done by the decimal point


Depending on the configuration (Using front keys: oP 063) Gross- and/or Net- and/or Tare weight is
filed, see chapter 4.3.3 . Display of the datasets using VNC see chapter 4.4.3 .
The alibi memory can be completely erased, if the CAL switch is open, using front keys (AL 101).

With the Browser the datasets can
be exported as CSV file for later
evaluation as EXCEL file.


The CSV file uses the 'comma' as separator between the different elements of an item.
Consequently in the EXCEL program a conversion has to be activated to receive only one element
per column.
The sequence number is preceded by an additional line number.

Example (Dataset as CSV-file):

Line,Seq.No.,Date,Time,WP,Type,Weight

55,13,2007-05-22,16:26:05,A,B,"12.450 kg"
56,13,2007-05-22,16:26:05,A,N,"10.400 kg"
57,13,2007-05-22,16:26:05,A,T,"2.050 kg"


After conversion in EXCEL:


Take care: EXCEL can import at maximum 65536 lines, if these are exceeded, the CSV-file has to
be divided previously.