Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 127
Section 6: PR1612 Commands
10/08 127
51207 Revision 1
6
PR 1612 Commands
This function enables the use of the following commands of the PR 1612 command set.
No license is required for using.
6.1 M
AIN COMMANDS FOR INDICATOR FUNCTION
Command
Reaction
Description
WGA
QGAwwwwwwemz
Read gross weight
WNA
QNAwwwwwwemz
Read net weight
WTA
QTAwwwwwwemz
Read tare weight
WDA
QDAwwwwwwemz
Read difference weight
wwwwwwe = Weight with sign and exponent
(e.g. '002340' = 002.34g; 001005 = 100kg)
m = no motion / Stand still,
z = 1: Tared, z = 2: Test, z = 4: Zero, 1/4d
WZA
Q
Set indicator to zero
WSA
Q
Set tare
WFA
Q
Reset tare
ZSC
QZSC
Reset communication
V
QV5410-1.00
Version request
Comman
d
Reaction
Description
I
Q
I
aaaaaa ddssnnnnn
LV
QLVxxxx/xx-x.xxaaaaaa ddssnnnnnttttt
xxxx/xx-x.xx = instrument version
aaaaaa = Max in 5 digits with dec. point,
dd = Unit, ss = Scale interval,
nnnnn = Span in mV/V
ttttt = deadload in mV/V
Befehl
Antwort
Beschreibung
WSAwwwwwwe
Q
Set fixtare
ZS1Awwwwwwewwwwwwe QZS1A
Set Limit 1
ZS2Awwwwwwewwwwwwe QZS2A
Set Limit 2
ZS3Awwwwwwewwwwwwe QZS3A
Set Limit 3
ZSDDDMMYYYYhhmm
QZSD
Set date and time
6.2 E
RROR
M
ESSAGES FOR
PR 1612
C
OMMANDS
Error
Description
Error
Description
E20000 Command unknown / failure
E10010 Arithmetic overflow
E30000 BCC error
E10020 Input > Max + Ovl
E50000 Analog test active
E10030 Input > 36 mV or ADU failure
E60000 Wrong address / weighing point
E10040 Weight exceeds display digits
E63000 Limit, fixtare > Max
E10050 Weight not available
E70000 Scale tared / in motion
E10060 No sense voltage
E10070 Negative measuring signal