Section 4: Commissioning
10/08 80
51207 Revision 1
4.5.10 Linearization
After selecting
-[Calibration]-[New]/[Modify] and after completing calibration, select the
linearization menu with softkey [Linear.]:
by load
by mV/V
by data
Linear.
Test
When you press [Linear.] the menu shown below appears:
WP A/Calibration/Linearisation
Max at
3000.00 kg
1.000000
Add
by mV/V
by load
To add a new linearization point, press [Add], fill in the weight value, apply the weight and press [by
load]. Then fill in the corresponding value in mV/V for the weight. After pressing [mV/V], the value
can be entered directly.
Up to 3 linearization points can be determined.
A linearization point can be added with [Add], removed with [Delete] and changed with [Change].
WP A/Calibration/Linearisation
1. Lin. point
750 kg
0.250010
2. Lin. point
1500 kg
0.500020
3. Lin. point
2250 kg
0.750040
Max at
3000.00 kg
1.000000
Add
Change
Delete
by mV/V
by load
A linearization point can be selected with
, changed with [Change] and deleted with [Delete].
4.5.11 Test Value Determination / Display
Press [Test] to activate test value calculation. The maximum capacity (Max) is displayed with the
designation
without a weight unit.
WP A/Calibration
Test
Depending on the settings under
-[Calibration]-[Param]-[Test mode], either is shown when you
press
to view the test data.
- the maximum capacity with [Absolute], or
- the deviation from the test value with [Relative]