Section 4: Commissioning

10/08 80

51207 Revision 1

4.5.10 Linearization

After selecting

-[Calibration]-[New]/[Modify] and after completing calibration, select the

linearization menu with softkey [Linear.]:

by load

by mV/V

by data

Linear.

Test



When you press [Linear.] the menu shown below appears:

WP A/Calibration/Linearisation

Max at

3000.00 kg

1.000000

Add

by mV/V

by load


To add a new linearization point, press [Add], fill in the weight value, apply the weight and press [by
load]. Then fill in the corresponding value in mV/V for the weight. After pressing [mV/V], the value
can be entered directly.
Up to 3 linearization points can be determined.
A linearization point can be added with [Add], removed with [Delete] and changed with [Change].

WP A/Calibration/Linearisation

1. Lin. point

750 kg

0.250010

2. Lin. point

1500 kg

0.500020

3. Lin. point

2250 kg

0.750040

Max at

3000.00 kg

1.000000

Add

Change

Delete

by mV/V

by load

A linearization point can be selected with

, changed with [Change] and deleted with [Delete].

4.5.11 Test Value Determination / Display


Press [Test] to activate test value calculation. The maximum capacity (Max) is displayed with the
designation

without a weight unit.

WP A/Calibration

Test

Depending on the settings under

-[Calibration]-[Param]-[Test mode], either is shown when you

press

to view the test data.

- the maximum capacity with [Absolute], or
- the deviation from the test value with [Relative]