beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 31

background image

Section 3: Installing the Instrument and Plug-In Cards

10/08 31

51207 Revision 1

3.4.5 PR 5510/07 Analog Input/Output Card


The plug-in card with 15-contact D-Sub socket for the analog channels is configured as 1 analog
output (active) and 4 analog inputs. The analog inputs are not supported by the standard
instrument.

Internal
connection:

34-contact connector socket on flat
cable for Slot 1, 2

Number of outputs: 1 active current output: 20mA, 10V

output voltage via external 500 ohm
resistor

Output
assignment:

Gross / net / display following,
configurable

Output range:

0/4 ... 20mA, configurable

Output resolution: 16 bits binary,

20,000 internal counts @ 20 mA

Output linearity
error:

@ 0 - 20mA: 0.04 %
@ 4 - 20 mA: 0.02 %

Output
temperature effect:

< 100 ppm/K

Output zero error: 0.05 %
Max. output error: < 0.1 %

Load:

Max. 0 ... 500 ohms

External
connection:

15-contact D-Sub female connector

Accessories:

15-pin D-Sub male connector

Dimensions:
(LxWxH):

86x53x16mm

Weight:

40g

Inputs:

4 channels current or voltage input
0 ... 20 mA, input resistance
250 ohms
0 ... 10 V, input resistance
100 kohms
0 ... 5 V, input resistance >10
Mohms

Input resolution:

3,000 internal counts corresponding
to e.g. 0 - 20 mA / 0 ... 10 V

Input accuracy:

0.2 %

Input,
linearity error:

< 0.03 %

Input
temperature effect:

< 50 ppm/K

Input, reserve:

+- 15%, i.e. –1.5V ... +11.5V

Electrical isolation: Output: yes, inputs: no