Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 31
Section 3: Installing the Instrument and Plug-In Cards
10/08 31
51207 Revision 1
3.4.5 PR 5510/07 Analog Input/Output Card
The plug-in card with 15-contact D-Sub socket for the analog channels is configured as 1 analog
output (active) and 4 analog inputs. The analog inputs are not supported by the standard
instrument.
Internal
connection:
34-contact connector socket on flat
cable for Slot 1, 2
Number of outputs: 1 active current output: 20mA, 10V
output voltage via external 500 ohm
resistor
Output
assignment:
Gross / net / display following,
configurable
Output range:
0/4 ... 20mA, configurable
Output resolution: 16 bits binary,
20,000 internal counts @ 20 mA
Output linearity
error:
@ 0 - 20mA: 0.04 %
@ 4 - 20 mA: 0.02 %
Output
temperature effect:
< 100 ppm/K
Output zero error: 0.05 %
Max. output error: < 0.1 %
Load:
Max. 0 ... 500 ohms
External
connection:
15-contact D-Sub female connector
Accessories:
15-pin D-Sub male connector
Dimensions:
(LxWxH):
86x53x16mm
Weight:
40g
Inputs:
4 channels current or voltage input
0 ... 20 mA, input resistance
250 ohms
0 ... 10 V, input resistance
100 kohms
0 ... 5 V, input resistance >10
Mohms
Input resolution:
3,000 internal counts corresponding
to e.g. 0 - 20 mA / 0 ... 10 V
Input accuracy:
0.2 %
Input,
linearity error:
< 0.03 %
Input
temperature effect:
< 50 ppm/K
Input, reserve:
+- 15%, i.e. –1.5V ... +11.5V
Electrical isolation: Output: yes, inputs: no