Section 4: Commissioning

10/08 65

51207 Revision 1

4.3.5 Subsequent Change of Deadload using Front Keys


If due to mechanical modifications the deadload (weight of empty scale / hopper) has changed, the
procedure is as follows:

START
1. Warm-up the instrument (minimum 0,5 hours), open CAL switch

2. Press

, SEtuP is shown

3. Press

, Cd 000 (Cd is flashing), switch to parameter number with , (000 is flashing),

change with

to Cd 003

MODIFY DEADLOAD (Cd 003)
4. Press

, dEAdLo is shown

5. Press

, Press

, LoAd is shown (Deadload with empty scale)

6. Press

, unLoAd is shown, unload scale

7. Press

, a weight value for deadload is shown

8. Press

, deadload is set, the display must show 00000.0 kg

9. Press

, Cd 004

STORING and EXIT
10. Press

, SAVE is shown

11. Press

, Press

, YES is shown, (if values must not be saved, alter with

to no )

12. Press

, during storage time SAVE. . . is shown, when finished Cd 000,

mit

wird das Setup-Menü verlassen.

13. If no further calibration data or parameters have to be entered,

close the CAL switch now, see chapter 4.1.1 .


4.3.6 View Calibration Data

START

1. Press

, SEtuP is shown

2. Press

, Cd 000 (Cd is flashing)

3. Press

, CALIb is shown (with CAL switch closed CAL.CLS is shown, calibration is not possible,

but data can be shown, continue with 6.)

4. Press

, nEW is shown, with

switch to VIEW

5. Press

, CAL.oPn is shown for CAL switch open

6. Press

, MAX.FSd is shown (Maximum load = Full scale deflection)

7. Press

, the entered maximum load is shown

8. Press

, Cd 002

9. Press

, StEP is shown

10. Press

, the entered scale interval is shown

11. Press

, Cd 003

12. Press

, dEAdLo is shown

13. Press

, the deadload in weight is shown

14. Press

, dEAdLo is shown

15. Press

, the equivalent for deadload in mV/V is shown

16. Press

, Cd 004

17. Press

, SPAn is shown

18. Press

, the equivalent for maximum load in mV/V is shown

19. Press

, Cd 005

20. Press

, 0-rnG is shown

21. Press

, the already used zeroset range is shown

22. Press

, Cd 006

23. Press

, uVoLt.d is shown

24. Press

, the measuring signal in µV/d is shown.


The other parameters can be displayed in the same way.