Section 4: Commissioning

10/08 64

51207 Revision 1

4.3.4 New Calibration using Front Keys


Example: Max (Maximum load) 600 kg, scale interval 200 g, setting deadload with empty scale.
At start of a new calibration the instrument has got the following default values: Max 3000 kg, scale
interval 1 kg.

START
1. Warm-up the instrument (minimum 0,5 hours), open CAL switch

2. Press

, SEtuP is shown


NEW CALIBRATION (Cd 000)
3. Press

, Cd 000 (Cd is flashing)

4. Press

, CALIb is shown (with CAL switch closed CAL.CLS is shown, calibration is not possible,

but data can be shown)

5. Press

, nEW (for new calibration) is shown


DEFINING SCALE RANGE and WEIGHT UNIT (Cd 001)
6. Press

, Cd 001

7. Press

, MAX.FSd for Max (Maximum load for weight display = Full scale deflection) is shown

8. Press

, - - - - - -.kg is shown

9. With the decimal point is shifted left for 1 digit behind the decimal point - - - - -.- kg
10. Press

00300.0 kg is shown (the 3 is flashing) , with (3 x) change to 6 für 600.0 kg

11. With (4 x) switch to weight unit (kg is flashing) and if required change to t, Lb or g with

DEFINING SCALE INTERVAL (Cd 002)
12. Press

, Cd 002

13. Press

, StEP is shown

14. Press

, 1 is shown, change with to 2 for 200 g scale interval


SETTING DEADLOAD (Cd 003)
15. Press

, Cd 003

16. Press

, dEAdLo is shown

17. Press

, LoAd is shown (Deadload with empty scale, else with

MUoLt for mV/V input)

18. Press

, unLoAd is shown, unload scale

19. Press

, a weight value for deadload is shown

20. Press

, deadload is set, the display must show 00000.0 kg


SETTING SPAN (Cd 004)
21. Press

, Cd 004

22. Press

, SPAn is shown

23. Press

, LoAd is shown (Range with test weights, else with

MUoLt for mV/V input)

24. Press

and apply test weights, instrument displays a weight value (still not calibrated)

25. Press

, WEIGHt is shown

26. Press

, 00600.0 kg is shown (Maximum test weight),

alter with / und

/

the value for the test weight

27. Press

, Cd 005


STORING and EXIT
28. Press

, SAVE is shown

29. Press

, YES is shown, (if values must not be saved, alter with

to no )

30. Press

, during storage time SAVE. . . is shown, when finished Cd 000,

leave the setup menu with

31. If the calibration parameters CP 010 - CP 026 have not to be / are already set,

close the CAL switch now, see chapter 4.1.1 .