S2 0 1, S2 0 2 – Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 32
Section 3: Installing the Instrument and Plug-In Cards
10/08 32
51207 Revision 1
Analog input
Signal selection
Channel
Current
0...+ 20mA DC
Voltage
0...+ 10V DC
Voltage
0...+ 5V DC
S
2
0
1
CH1
CH2
CH3
CH4
----
ON
ON
ON
ON
----
OFF
OFF
OFF
OFF
----
OFF
OFF
OFF
OFF
----
S
2
0
2
CH1
CH2
CH3
CH4
----
OFF
OFF
OFF
OFF
----
ON
ON
ON
ON
----
OFF
OFF
OFF
OFF
----
Input resistance
250 ohms
100 kohms
> 10 Mohms
Pin allocation of rear socket (for installation, see Chapter 3.4.1)
D-Sub 15
Input (PIN)
2
9
3
10
4
11
5
12
4 channels
+ CH1 GND + CH2 GND + CH3 GND + CH4 GND
Output (PIN)
13
14
1 channel
+ 0/4 ... 20
mA
Sig GND