Section 4: Commissioning
10/08 87
51207 Revision 1
[Remote display]
Select the serial interface to which the remote display is connected and then select [Param] to
define the [Baudrate] and the remote display type [Mode] connected.
Setup/Serial ports/Builtin RS232
Assigned to
Remote display
Protocol
Remote display
Baudrate
9600 bd Select [Baudrate] with
and
Bits
7
set the baud rate with
Parity
even
Stopbits
1
Mode
single transmitter
The available baud rates are 300, 600, 1200, 2400, 4800 or 9600. The parameters (displayed in
light-gray) cannot be changed.
If only 1 instrument is connected to a PR1627/8 (normal case), [Mode] must be set to [single
transmitter].
Foe a PR1577 remote display [PR 1577 mode] has to be set.
If more than 1 instrument are connected to 1 remote display, [Mode] must be set to [multiple
transmitters].
At [Device Id] the own instrument address (A, B, C ...) has to be entered, at [Next Device Id] the
address of the subsequent instrument has to be entered.
[JBUS/MOD-Bus]
Setup/Serial ports/Builtin RS232
Assigned to
JBUS/MOD-Bus
Protocol
JBUS/MOD-Bus
Baudrate
9600 bd Select [Baudrate] with
and
Bits
8
set the baud rate with
Parity
even None, odd or even
Stopbits
1
Slave
1 Select an address; 1 to 255
Press
to return to the [Serial ports parameter] setup menu. When you close this menu, the
following messages are displayed if at least one setting was changed:
Save settings
Start JBUS/MOD-Bus
The JBUS/MOD bus protocol is described in Chapter 4.9.