beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 96

background image

Section 4: Commissioning

10/08 96

51207 Revision 1

2. Determining a condition

The selected action of each digital input can be combined with a condition that must be met for
signal change from 0 to 1 (on) or for signal change from 1 to 0 (off). Select the condition from the list;
see Chapter 4.8.1. No condition is defined when selecting [no condition]; the action is executed
directly.

Example: Taring via digital input, only if the instrument is set to gross:

Setup/Digital I/O parameter

Output 1

marker bit 1

X64=1

Output 2

marker bit 2

X65=1

Output 3

marker bit 3

X66=1

Input 1 on

set tare

X113=1

Condition

tare active

X58=0

Input 1 off

-no action-

Input 2 on

-no action-

Input 2 off

-no action-

Input 3 on

-no action-

Input 3 off

-no action-

BCD out

Gross


When input 1 changes from 0 to 1 [Input 1 on], a taring signal is activated only, if the condition under
[Condition] is met (here: the scale was not tared previously; this is determined by [tare active X58=0]).