background image

Section 4: Commissioning

10/08 61

51207 Revision 1

4.3 C

ONFIGURATION AND

C

ALIBRATION USING

F

RONT

K

EYS

4.3.1 Meaning of Front Keys


The instrument can be set directly via front keys:


Press Setup

SETUP w

Setup-Menu, press

CD 000w

Calibration menu, 'Cd' is flashing


With

the next parameter group (Cp) is selected, with switch to the parameter number ('000' is

flashing), the parameter table is described in chapter 4.3.3 .

If a selection has to be performed,

/

have to be used.

The digit position is selected by / , with the digit is increased, with decreased.

Inputs are completed by

, a menu is left with

.


All parameters / data are clearly characterized by a max. 3-digit number, additionally two letters are
preceded for the parameter group for easy allocation.

If data have to be overwritten, a safety prompt appears:

SAVE w

if the data should not be saved, then 'YES' has the be changed with

to 'no'.


4.3.2 Entering Date and Time

1. Press

, SEtuP is shown

2. Press

, Cd 000 (Cd is flashing),

with

(9 x) change to dt 086

3. Press

, YEAr is shown

4. Press

, 2007 (YYYY) is shown, alter with / and

/

5. Press

, dAtE is shown

6. Press

, 05.15 (MM.DD) is shown, alter with / and

/

7. Press

, tIME is shown

8. Press

, 17.35 (HH.MM) is shown, alter with / and

/

9. Press

, dt 086 is shown

10. Press

to leave the menu