Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 66
Section 4: Commissioning
10/08 66
51207 Revision 1
4.3.7 View Calibration Data for Deadload and Maximum Load
Normally all settings of the instrument should be documented or printed out during commissioning.
If an instrument has to be replaced by another one due to breakdown and no new calibration can /
should be done, the values for deadload and maximum load in mV/V are absolutely required.
Therefore the following has to be done:
1. Press
, SEtuP is shown
2. Press
, Cd 000 (Cd is flashing), with switch to parameter number, (000 is flashing),
with
(3 x) change to Cd 003
3. Press
, dEAdLo is shown
4. Press
, LoAd is shown, with
change to MVoLt
5. Press
, the deadload is shown in mV/V
6. Note the value for deadload in mV/V
7. Press
, Cd 004 is shown
8. Press
, SPAn is shown
9. Press
, MVoLt is shown
10. Press
, the maximum load is shown in mV/V
11. Note the value for maximum load in mV/V
12. Leave Setup menu
4.3.8 PIN Code
The Setup-program can be protected against unauthorized access by a PIN-code defined by the
user.
ACTIVATING PIN
1. Press
, SEtuP is shown
2. Press
, Cd 000 (Cd is flashing)
3. Select oP with
(6 x), oP 060 is shown
4. With switch to parameter number and with
switch to oP 061
5. Press
, PIn is shown
6. Press
, the default PIN (000000 = no request) is shown,
it can be changed with / and
/
7. Press
, PIn is stored and oP 062 is shown
8. The menu can be left with
and
.
ENTERING PIN
1. Press
, SEtuP is shown
2. Press
, PIn is shown
3. Press
, 000000 is shown, enter PIN with / and
/
4. Press
, with wrong PIN PInWro is shown, back to 3.
with correct PIN Cd 000 (Cd is flashing) is shown.
SUPER-PIN
If the PIN-Code is lost, the Setup can be unlocked with Super-PIN '212223'.