beautypg.com

Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 81

background image

Section 4: Commissioning

10/08 81

51207 Revision 1

4.5.12 Finishing / Saving the Calibration

Finish the calibration with

.

You are prompted to confirm whether calibration should be closed
without determining the test value.

Calibration not complete

Exit calibration?

If not all data was determined when calibrating with [New] (e.g. dead load
not set/entered), this message is shown:

Press [Yes] to confirm and then press

again; another prompt is

displayed:

Exit calibration ?

Press [Save] to save changes in calibration data. If you press [Undo],
changes are not saved and the display returns to the selection menu for
the weighing points.

Saving calibration

Confirm, or

Exit calibration

close the menu.


After finishing the calibration, set the CAL switch to the closed position; see also Chapter 4.1.1.

4.5.13 Parameter Input

Open the menu via

-[Calibration]-[Param]

WP A/Calibration

Measuretime

160 ms

Digital filter

bessel

Fcut

3.00 Hz This line is shown only, if the filter is on.

Test mode

absolute

W & M

none

Standstill time

0.50 s

Standstill range

1.00 d


[Measuretime]
Select the measuretime; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms. In
'legal-for-trade' mode select <= 1 s.

[Digital filter]
The digital filter can be switched on only with the measuretime set to <= 160 ms.
Select the filter characteristic: [off], [bessel], [aperiod.], [butterw.], [tscheby.].

[Fcut]
Enter the cut-off frequency for the low-pass filter (0.1 – 80 Hz); the setting is dependent on the
measuretime.