Fairbanks X SERIES PR 5410 Process Indicator User Manual
Page 92
Section 4: Commissioning
10/08 92
51207 Revision 1
4.7.8 Network Parameters
You can configure settings for the network connections (built-in LAN adapter) under
-[Network
parameter].
Setup/Network parameter
HW address
00:90:6C:FB:E9:85
Fixed address determined by the instrument
Hostname
PR5410-UDOWEIGH Device name; user-definable*
Use DHCP
Address is assigned by the server
IP Address
172.24.21.82 Address assigned by the server
Subnetmask
255.255.240.0 Mask for permissible IP address range
Standardgateway
0.0.0.0 IP number for gateway
Remote access
VNC-Client
255.255.255.255
Permissible client for operation of the
instrument
*The [Hostname] is subject to the following restrictions:
Minimum number of characters: 2, maximum number of characters: 24
The first character must be a letter. Spaces are not permitted.
0-9, A-Z (upper and lower case letters are not distinguished) are permitted.
- or . may be included, but neither at the end nor in succession.
When setting [IP address], [Subnet mask] and [Standardgateway], please consult with your network
administrator.
You can configure access permissions for the address specified under [VNC client]:
VNC client
0.0.0.0. Access over VNC not permitted
VNC client
172.24.21.101 Access only from client machine with this address
VNC client
172.24.21.255 Access from any client with address within range 172.24.21.1 to
254
VNC client
255.255.255.255 Access from client with any address