Table 0-15: Control Input Pin Assignments

Input

Status

Condition when enabled

A

EXTERNAL ZERO
CAL

Zero calibration mode is activated. The mode field of the
display will read ZERO CAL R.

B

EXTERNAL SPAN

CAL

Span calibration mode is activated. The mode field of the

display will read SPAN CAL R.

C

EXTERNAL LOW

SPAN CAL

Low span (mid-point) calibration mode is activated. The

mode field of the display will read LO CAL R.

D

Unused

E

Unused

F

Unused

DIGITAL GROUND Provided to ground an external device (e.g., recorder).

U

DC power for
Input pull ups

Input for +5 VDC required to activate inputs A - F. This
voltage can be taken from an external source or from the

“+” pin.

+

Internal +5V
Supply

Internal source of +5V which can be used to activate
inputs when connected to pin U.

There are two methods to activate control inputs. The internal +5V available from the “+”

pin is the most convenient method (see figure). However, to ensure that these inputs are

truly isolated, a separate, external 5 VDC power supply should be used.

LOW SPAN

ZERO CA

L

SPAN C

A

L

CONTROL IN

Local Power Connections

External Power Connections

LOW SPAN

ZERO CA

L

SPAN C

A

L

CONTROL IN

-

+

5 VDC Power

Supply

A B C D E F U

+

A B C D E F U

+

Figure 0-5: Control Inputs with Local and External 5 V Power Supply

6.9. Setup - Communication Ports (COMM)

The M9110E is equipped with two serial communication ports located on the rear panel

(Figure 3-3). Both ports operate similarly and give the user the ability to communicate with,

issue commands to, and receive data from the analyzer through an external computer
system or terminal. By default, both ports operate on the RS-232 protocol. The COM2 port,