Single range mode (sngl), Independent range mode (ind) – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 68
Operating Instructions
Model 9110E Instruction Manual
54
M9110E Rev B
6.5.3. Single Range mode (SNGL)
The default range mode for the analyzer is single range, in which all analog concentration
outputs (usually A1, A2 and A3) are set to the same reporting range. This reporting range
can be set to any value between 100 ppb and 20000 ppb. However, the electronic signal
ranges of the analog outputs may still be configured for different values (e.g., 0-10 V and
0-0.1 V).
To select SNGLE range mode and to set the upper limit of the range, press:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP C.3
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE MODE: SNGL
SNGL
DUAL AUTO
ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP C.3
RANGE MODE: SNGL
SNGL
DUAL AUTO
ENTR EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
NOTE
If a value <10.0
or >20.000 is
entered, the
ENTR key will
disappear
6.5.4. Independent Range Mode (IND)
The independent range mode allows the concentration outputs A1, A2 and A3 to be
configured with a different range value. In IND range mode the RANGE test function
displayed on the front panel will then be replaced by three separate functions:
• RANGE1: Range value for output A1 (NO
x
), e.g., 0-10 V for 0-1500 ppm NO
X
.
• RANGE2: Range value for output A2 (NO), e.g., 0-10 V for 0-1000 ppm NO.
• RANGE3: Range value for output A3 (NO
2
), e.g., 0-10 V for 0-500 ppm NO
2
.