Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 252
APPENDIX A
Model 9110E Instruction Manual
238
044100120 Rev B
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
SG_NO_DWELL
Seconds
1
0.1–30
Dwell time after switching valve to
NO position in single-gas measure
modes.
NO_SAMPLE
Samples
2
1-30
Number of samples to take in NO
mode.
SG_NO_SAMPLE
Samples
1
1–30
Number of samples to take in NO
mode in single-gas measure
modes.
USER_UNITS — PPB PPB
PPM
UGM
MGM
Concentration units for user
interface. Enclose value in double
quotes (") when setting from the
RS-232 interface.
DIL_FACTOR
—
1
1-1000
Dilution factor. Used only if is
dilution enabled with
FACTORY_OPT
variable.
AZERO_ENABLE
—
ON
ON, OFF
ON enables auto-zero; OFF
disables it.
AZERO_FREQ Minutes
1
0-60
Auto-zero
frequency.
AZERO_DWELL
Seconds
2
0-60
Dwell time after opening or closing
auto-zero valve.
AZERO_SAMPLE Samples 2
1-10 Number of auto-zero samples to
average.
SG_AZERO_SAMP
Samples
2
1–10
Number of auto-zero samples to
average in single-gas measure
modes.
AZERO_FSIZE Samples
15
1-50
Moving
average filter size for auto-
zero samples.
AZERO_LIMIT
mV
200
0-1000
Maximum auto-zero offset allowed.
NOX_SPAN1 Conc.
400 4-20000
Target
NO
X
concentration during
span calibration of range 1.
NO_SPAN1
Conc.
400
4-20000
Target NO concentration during
span calibration of range 1.
NO2_SPAN1 Conc.
400 4-20000
Target
NO
2
concentration during
converter efficiency calibration of
range 1.
NOX_SLOPE1 PPB/mV
3,5
PPM/mV
4
1 0.25-4
NO
X
slope for range 1.
NOX_OFFSET1 mV
0
-10000-
10000
NO
X
offset for range 1.
NO_SLOPE1 PPB/mV
3,5
PPM/mV
4
1
0.25-4
NO slope for range 1.
NO_OFFSET1 mV 0
-10000-
10000
NO offset for range 1.
NOX_SPAN2 Conc.
400 4-20000
Target
NO
X
concentration during
span calibration of range 2.
NO_SPAN2
Conc.
400
4-20000
Target NO concentration during
span calibration of range 2.