APPENDIX A

Model 9110E Instruction Manual

248

044100120 Rev B

Signal Name

Bit or Channel

Number

Description

ST_SPAN_CAL

4

0 = in span calibration

1 = not in span

ST_DIAG_MODE

5

0 = in diagnostic mode

1 = not in diagnostic mode

6-7

Spare

B status outputs, U27, J1018, pins 1-8 = bits 0-7, default I/O address 324 hex

0-7

Spare

Front panel I

2

C keyboard, default I

2

C address 4E hex

MAINT_MODE

5 (input)

0 = maintenance mode

1 = normal mode

LANG2_SELECT

6 (input)

0 = select second language
1 = select first language (English)

SAMPLE_LED

8 (output)

0 = sample LED on
1 = off

CAL_LED

9 (output)

0 = cal. LED on
1 = off

FAULT_LED

10 (output)

0 = fault LED on

1 = off

AUDIBLE_BEEPER

14 (output)

0 = beeper on (for diagnostic testing only)
1 = off

Relay board digital output (PCF8575), default I

2

C address 44 hex

RELAY_WATCHDOG 0

Alternate

between

0 and 1 at least every 5 seconds

to keep relay board active

RCELL_HEATER

1

0 = reaction cell heater on
1 = off

CONV_HEATER

2

0 = converter heater on
1 = off

MANIFOLD_HEATER

2

3

0 = bypass or dilution manifold heater on
1 = off

IZS_HEATER

4

0 = IZS heater on

1 = off

5

Spare

SPAN_VALVE

6

0 = let span gas in
1 = let zero gas in

CAL_VALVE

7

0 = let cal. gas in

1 = let sample gas in

AUTO_ZERO_VALVE

8

0 = let zero air in

1 = let sample gas in

NOX_VALVE

9

0 = let NO

X

gas into reaction cell

1 = let NO gas into reaction cell

10-15

Spare

Rear board primary MUX analog inputs, MUX default I/O address 32A hex

PMT_SIGNAL 0

PMT

detector

HVPS_VOLTAGE

1

HV power supply output

PMT_TEMP 2

PMT

temperature

3

Spare