Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 254
APPENDIX A
Model 9110E Instruction Manual
240
044100120 Rev B
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
RCELL_SAMP_RATIO
—
0.53
0.1-2
Maximum reaction cell pressure /
sample pressure ratio for valid
sample flow calculation.
STD_BOX_TEMP ºK
298
Valid limits:
278-338
1-500
Standard box temperature and
valid limits for temperature
compensation.
STD_RCELL_TEMP ºK
323
Valid limits:
278-338
1-500
Standard reaction cell temperature
and valid limits for temperature
compensation.
STD_RCELL_PRESS "Hg
7
Valid limits:
0.5-12
0.1-50
Standard reaction cell pressure
and valid limits for pressure
compensation.
STD_SAMP_PRESS "Hg
29.92
Valid limits:
0.5-32
0.1-50
Standard sample pressure and
valid limits for pressure
compensation.
PRESS_FILT_SIZE
5
Samples
60
1-120
Sample and reaction cell pressure
moving average filter size.
PRESS_SAMP_FREQ
5
Minutes
2
0.1-60
Sample and reaction cell pressure
sampling frequency.
RS232_MODE
—
0
0-65535
RS-232 COM1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable iChip support
16 = enable Hessen protocol
32 = enable multidrop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF support
1024 = enable RS-485 mode
2048=even parity, 7 data bits, 1
stop bit
4096 = enable command prompt
BAUD_RATE — 19200
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate. Enclose
value in double quotes (") when
setting from the RS-232 interface.
MODEM_INIT
—
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0 E0
Q1 &W0”
Any
character
in the
allowed
character
set. Up
to 100
character
s long.
RS-232 COM1 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually. Enclose value in double
quotes (") when setting from the
RS-232 interface.
RS232_MODE2
BitFlag
11
0-65535
RS-232 COM2 mode flags.
(Same settings as RS232_MODE)