Operating Instructions

Model 9110E Instruction Manual

60

M9110E Rev B

To access and navigate the VARS menu, use the following key sequence:

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=15.0 Minutes

NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

1 ) TPC_ENABLE=ON

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

4) SFLOW_SET=500.0 cc/m

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

ENTER VARS PASS: 818

8 1 8

ENTR EXIT

SETUP X.X

DAS_HOLD_OFF=15.0 Minutes


1 5 .0

ENTR EXIT

SETUP X.X

5) OFLOW_SET=

80.0 cc/m

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

6) IZS_SET=80.0 cc/m

PREV NEXT JUMP

EDIT PRNT EXIT

DO NOT change

theses set-points

unless

specifically

instructed to by

T-API Customer

Service.

SETUP X.X

CLOCK_ADJ=0 Sec/Day


+ 0 0

ENTR EXIT

SETUP X.X

7) CLOCK_ADJ=0 Sec/Day

PREV NEXT JUMP

EDIT PRNT EXIT

EXIT ignores the new setting.

ENTR accepts the new setting.

SETUP X.X

1 ) TPC_ENABLE=ON


ON

ENTR EXIT

SETUP X.X

2 ) DYN_ZERO=ON

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

2 ) DYN_ZERO=ON


ON

ENTR EXIT

Toggle this keys to change setting

Toggle this keys to change setting

SETUP X.X

3) DYN_SPAN=ON

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

3 ) DYN_SPAN=ON


ON

ENTR EXIT

Toggle this keys to change setting

Toggle this keys to change setting

SETUP X.X

8) CAL_ON_NO

2

= OFF

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

8) CAL_ON_NO

2

= ON


ON

ENTR EXIT

Toggle this keys to change setting

Toggle this keys to change setting