6.7.8. Flow Calibration

The flow calibration allows the user to adjust the values of the sample and ozone flow rates

as they are displayed on the front panel and reported through COM ports to match the

actual flow rate measured at the respective pneumatic ports. This does not change the
hardware measurement of the flow sensors, only the software calculated values. This is one

of the two parameters in the DIAG menu that is retained after you exit the menu. To carry
out this adjustment, connect an external, sufficiently accurate flow meter to the respective

test port (Chapter 11 contains more details on this setup). Once the flow meter is attached

and measures the actual flow, follow this key sequence:



SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG ACAL DAS RNGE PASS CLK MORE EXIT

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

Exit at

any time
to return

to main

the

SETUP

menu

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Exit returns

to the

previous menu

Repeat Pressing NEXT until . .

.

DIAG FLOW CALIBRATION

PREV NEXT

ENTR EXIT

DIAG FCAL ACTUAL FLOW: 480 CC / M

0 4 8 0 ENTR EXIT

ENTR accepts the

new value and

returns to the

previous menu

EXIT ignores the

new value and

returns to the

previous menu

Adjust these values until

the displayed flow rate

equals the flow rate being

measured by the

independent flow meter.

DIAG FLOW SENSOR TO CAL: SAMPLE

SAMPLE OZONE

ENTR EXIT

Choose between

sample and ozone

flow sensors.

SETUP X.X ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

6.8. Digital Inputs and Outputs

6.8.1. Status Outputs

The status output signals report analyzer conditions through optically isolated transistors

which can sink up to 50 mA of DC current to ground. These outputs can be used with

interface devices that accept logic-level digital inputs, such as programmable logic

controllers (PLC).

NOTE

Most PLC devices have internal provisions for limiting the current that the output

can draw from an external device. When connecting the M9110E status outputs to

a unit that does not have this feature, a 120 Ω external pull-up resistors must be

used to limit the current through the transistor output to 50 mA or less. Refer to

the motherboard schematic 04069 in Appendix D.