Model 9110E Instruction
Manual

Warranty/Repair

Questionnaire

Model 9110E

TELEDYNE API CUSTOMER SERVICE

EMAIL: [email protected]

PHONE: 626-934-1673

FAX: (626) 961-2538

(04503)044100130 Rev C

263

TELEDYNE

INSTRUMENTS

Advanced Pollution Instrumentation

A Teledyne Technologies Company

APPENDIX C: Warranty/Repair Questionnaire Model 9110E

CUSTOMER: ____________________________________ PHONE: ___________________________________________

CONTACT NAME: _______________________________ FAX NO. ___________________________________________

SITE ADDRESS: ______________________________________________________________________________________

MODEL 9110E SERIAL NO.: ________________________FIRMWARE REVISION: _______________________________

1. ARE THERE ANY FAILURE MESSAGES? ____________________________________________________________

_____________________________________________________________________________________________________

_____________________________________________________________________________________________________

PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT
)
*IF OPTION IS INSTALLED

PARAMETER RECORDED

VALUE

ACCEPTABLE

VALUE

RANGE

PPB/PPM

50 PPB TO 20 PPM

NOx STAB

PPB/PPM

≤ 1 PPB WITH ZERO AIR

SAMPLE FLOW

CM

3

500 ± 50

OZONE FLOW

CM

3

80 ± 15

PMT SIGNAL WITH ZERO

AIR

MV -20

TO

150

PMT SIGNAL AT SPAN GAS

CONC

MV

PPB

0-5000MV

0-20,000 PPB

NORM PMT SIGNAL AT

SPAN GAS CONC

MV

PPB

0-5000MV

0-20000PPB

AZERO

MV -20

TO

150

HVPS

V 400

900

RCELL TEMP

ºC

50 ± 1

BOX TEMP

ºC

AMBIENT ± 5ºC

PMT TEMP

ºC

7 ± 2ºC

IZS TEMP*

ºC

50 ± 1ºC

MOLY TEMP

ºC

315 ± 5ºC

RCEL PRESS

IN-HG-A <10

SAMP PRESS

IN-HG-A AMBIENT

±

1

NOx SLOPE

1.0 ± 0.3

NOx OFFSET

50 TO 150

NO SLOPE

1.0 ± 0.3

NO OFFSET

50 TO 150

ETEST

PMT MV

2000 ± 1000

OTEST

PMT MV

2000 ± 1000

Values are in the Signal I/O

REF_4096_MV

MV

4096mv ±2mv and Must be

Stable

REF_GND

MV

0± 0.5 and Must be Stable