beautypg.com

Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 231

background image

Model 9110E Instruction Manual

Troubleshooting & Repair

M9110E Rev B

217

11.5.15.1. Reaction Cell Pressure

Measure the voltage across test points TP1 and TP5. With the sample pump disconnected or

turned off, the voltage should be 4500 ± 250 mV. With the pump running, it should be 800-
1700 mV depending on the performance of the vacuum pump. The lower the reaction cell

pressure, the lower the resulting voltage is. If this voltage is significantly different, the
pressure transducer S1 or the board may be faulty. If this voltage is between 2 and 5 V, the

pump may not be performing well, check that the reaction cell pressure is less than
10 in-Hg-A (at sea level). Ensure that the tubing is connected to the upper port, which is

closer to the sensor’s contacts; the lower port does not measure pressure.

11.5.15.2. Sample Pressure

Measure the voltage across test points TP1 and TP4. With the sample pump disconnected or

turned off, this voltage should be 4500 ± 250 mV. With the pump running, it should be
about 0.2 V less as the sample pressure drops by about 1 in-Hg-A below ambient pressure.

If this voltage is significantly different, the pressure transducer S2 or the board may be
faulty. A leak in the sample system to vacuum may also cause this voltage to be between

about 0.6 and 4.5. Make sure that the front panel reading of the sample pressure is at
about 1 in-Hg-A less than ambient pressure. Ensure that the tubing is connected to the

upper port, which is closer to the sensor’s contacts; the lower port does not measure
pressure.

Figure 11-5: Pressure / Flow Sensor Assembly