Operating Instructions

Model 9110E Instruction Manual

52

M9110E Rev B

In order to compensate for CPU clocks which run fast or slow, there is a variable to speed

up or slow down the clock by a fixed amount every day. To change this variable, press:

SAMPLE RANGE = 500.0 PPB NOX =XXX.X


< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS DIAG

EXIT

3x EXIT returns

to the main SAMPLE display

SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes

NEXT JUMP

EDIT PRNT EXIT

Enter sign and number of seconds per

day the clock gains (-) or loses (+).

SETUPX.X 1 )TPC_ENABLE= ON

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X 8) CLOCK_ADJ=0 Sec/Day

PREV JUMP

EDIT PRNT EXIT

SETUP X.X

CLOCK_ADJ:0 Sec/Day


+ 0 0

ENTR EXIT

SETUP X.X

8) CLOCK_ADJ=0 Sec/Day


PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

ENTER VARS PASS: 818


8 1 8

ENTR EXIT

Continue to press NEXT until …

6.5. Setup - Range Configuration (RNGE)

The analyzer has four analog output signals, accessible through a connector on the rear

panel.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

Figure 0-1: Analog Output Connector Key

All of these outputs can be configured either at the factory or by the user for full scale
outputs of 0.1 V, 1V, 5V or 10V. Additionally A1, A2 andA3 may be equipped with optional

0-20 mA current loop drivers and configured for any current output within that range
(e.g. 0-20, 2-20, 4-20, etc.).

Channels A1, A2 and A3 report analog signals that are proportional to the NO

x

, NO and

NO

2

concentrations of the sample gas, respectively. Analog output voltage or current can be

scaled independently. Here is one configuration example for independent ranges:

A1 OUTPUT (NO

x

): Output Signal = 0-5 V