Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 261
Model 9110E Instruction Manual
APPENDIX A
044100110 Rev B
247
APPENDIX A-4:
M9110E Signal I/O Definitions, Revision C.8
Table A-4: M9110E Signal I/O Definitions, Revision C.8
Signal Name
Bit or Channel
Number
Description
Internal inputs, U7, J108, pins 9-16 = bits 0-7, default I/O address 322 hex
0-7
Spare
Internal outputs, U8, J108, pins 1-8 = bits 0-7, default I/O address 322 hex
ELEC_TEST
0
1 = electrical test on
0 = off
OPTIC_TEST
1
1 = optic test on
0 = off
PREAMP_RANGE_HI
2
1 = select high preamp range
0 = select low range
O3GEN_STATUS
3
0 = ozone generator on
1 = off
4-5
Spare
I2C_RESET
6
1 = reset I
2
C peripherals
0 = normal
I2C_DRV_RST
7
0 = hardware reset 8584 chip
1 = normal
Control inputs, U11, J1004, pins 1-6 = bits 0-5, default I/O address 321 hex
EXT_ZERO_CAL
0
0 = go into zero calibration
1 = exit zero calibration
EXT_SPAN_CAL
1
0 = go into span calibration
1 = exit span calibration
2-5
Spare
6-7
Always
1
Control inputs, U14, J1006, pins 1-6 = bits 0-5, default I/O address 325 hex
0-5
Spare
6-7
Always
1
Control outputs, U17, J1008, pins 1-8 = bits 0-7, default I/O address 321 hex
0-7
Spare
Control outputs, U21, J1008, pins 9-12 = bits 0-3, default I/O address 325 hex
0-3
Spare
Alarm outputs, U21, J1009, pins 1-12 = bits 4-7, default I/O address 325 hex
4-7
Spare
A status outputs, U24, J1017, pins 1-8 = bits 0-7, default I/O address 323 hex
ST_SYSTEM_OK
0
0 = system OK
1 = any alarm condition
ST_CONC_VALID
1
0 = conc. valid
1 = hold off or other conditions
ST_HIGH_RANGE
2
0 = high auto-range in use
1 = low auto-range
ST_ZERO_CAL
3
0 = in zero calibration
1 = not in zero