Model 9110E Instruction Manual

APPENDIX A

044100110 Rev B

247

APPENDIX A-4:

M9110E Signal I/O Definitions, Revision C.8

Table A-4: M9110E Signal I/O Definitions, Revision C.8

Signal Name

Bit or Channel

Number

Description

Internal inputs, U7, J108, pins 9-16 = bits 0-7, default I/O address 322 hex

0-7

Spare

Internal outputs, U8, J108, pins 1-8 = bits 0-7, default I/O address 322 hex

ELEC_TEST

0

1 = electrical test on

0 = off

OPTIC_TEST

1

1 = optic test on
0 = off

PREAMP_RANGE_HI

2

1 = select high preamp range
0 = select low range

O3GEN_STATUS

3

0 = ozone generator on
1 = off

4-5

Spare

I2C_RESET

6

1 = reset I

2

C peripherals

0 = normal

I2C_DRV_RST

7

0 = hardware reset 8584 chip
1 = normal

Control inputs, U11, J1004, pins 1-6 = bits 0-5, default I/O address 321 hex

EXT_ZERO_CAL

0

0 = go into zero calibration
1 = exit zero calibration

EXT_SPAN_CAL

1

0 = go into span calibration

1 = exit span calibration

2-5

Spare

6-7

Always

1

Control inputs, U14, J1006, pins 1-6 = bits 0-5, default I/O address 325 hex

0-5

Spare

6-7

Always

1

Control outputs, U17, J1008, pins 1-8 = bits 0-7, default I/O address 321 hex

0-7

Spare

Control outputs, U21, J1008, pins 9-12 = bits 0-3, default I/O address 325 hex

0-3

Spare

Alarm outputs, U21, J1009, pins 1-12 = bits 4-7, default I/O address 325 hex

4-7

Spare

A status outputs, U24, J1017, pins 1-8 = bits 0-7, default I/O address 323 hex

ST_SYSTEM_OK

0

0 = system OK
1 = any alarm condition

ST_CONC_VALID

1

0 = conc. valid

1 = hold off or other conditions

ST_HIGH_RANGE

2

0 = high auto-range in use
1 = low auto-range

ST_ZERO_CAL

3

0 = in zero calibration

1 = not in zero